Custom project

Olefins extraction system

A novel method for parallel extraction of plasticizers and stabilisers from polyolefins for HPLC, GC and TLC analysis.
Labman custom system: Olefins extraction system

Key features

Reflux ShakingReflux shaking at 190degreesC +/-2degreesC and 300 rpm +/-1rpm
Reaction VesselsReaction vessels hold 100ml of fluid
Liquid HandlingLiquid handling accuracies better than 0.5% CV
Post-Filtration and Pre-Filtration0.45um post-filtration and 100um pre-filtration
GC VialsPrepared sample is collected in GC vials
HPLC PreparationHPLC Preparation

Overview

System Overview.

A novel method for parallel extraction of plasticizers and stabilisers from polyolefins for HPLC, GC and TLC analysis. Samples of plastics in small dissolution reaction vessels have various solvents and reagents added to them. The reaction vessels are heated up to 190degreesC and mixed at 300rpm. An automated reflux lid reduces evaporation. After dissolution further reagent is added and an aliquot is removed from the gelatinous precipitation/solvent mixture through a PTFE frit. The aliquot is dispensed through a 0.45um filter, into GC vials.